Medical air compressor systems are governed by strict international standards (such as NFPA 99, ISO 7396-1, HTM 02-01, and ISO 8573-1 Class 1) to guarantee patient safety and continuous delivery.
Below is a breakdown of the technical specifications and standards for a medical air compressor system.
Medical compressed air must strictly meet gas purity standards (often categorized as Grade D Breathing Air or medical-grade air):
Oil-Free Guarantee: 100% oil-free design (0 mg/m oil content).
Dew Point / Moisture Control:
Desiccant / Regenerative Dryers: Pressure dew point down to -40C (-40F) to prevent moisture condensation and microbial growth in distribution lines.
Contaminant & Gas Limits (per NFPA 99 / European Pharmacopoeia):
Carbon Monoxide (CO): < 10
Carbon Dioxide < 500text
Oxygen Concentration: 20.4 - 21.4
Particulates: Filtration down to 0.01} 0.1 mutext{m}$ (Efficiency 99.999
| Parameter | Standard / Typical Value |
| Delivery Pressure | $50 text{ to } 55 text{ psig}$ ($3.5 text{ to } 3.8 text{ bar}$) at point of use |
| System Operating Pressure | $85 text{ to } 105 text{ psig}$ ($5.8 text{ to } 7.2 text{ bar}$) |
| Compressor Technologies | Oil-less Reciprocating / Piston (Teflon-sealed), Oil-free Rotary Scroll, or Oil-free Rotary Screw |
| Redundancy Setup | Duplex, Triplex, or Quadruplex with automatic sequence lead/lag control. System must carry full load with one unit out of service ($N+1$ redundancy). |
| Noise Level | $< 65 text{ to } 70text{ dBA}$ (sound-attenuated acoustic enclosures) |
| Power Supply | 3-Phase, 208230V / 460V, 60Hz (or 380415V, 50Hz) connected to essential hospital electrical emergency power systems. |
Air Intake: Located outdoors, elevated, away from engine exhausts, vents, and sources of pollution.
Receivers / Air Storage Tank:
ASME Code Section VIII stamped vertical or horizontal pressure vessel.
Internal corrosion-resistant coating (epoxy or galvanized) with automatic/solenoid condensate drain valves.
Purification & Filtration Package:
Pre-filters (particle separation)
Coalescing filters (liquid/oil aerosol removal)
Dual desiccant air dryers with auto-regeneration (100% redundant standby)
Activated carbon filters (odor/hydrocarbon removal)
Bacteria / HEPA final filters
Monitoring & Control Panel:
UL 508A listed control panel
Automatic lead/lag motor switching and hour-run meters
Continuous CO monitoring, High Moisture/Dew Point alarms, and Low Pressure alarms integrated into hospital BMS (Building Management System)

Price:
Price 42000 INR / Piece
Minimum Order Quantity : 1 Piece
Mute : No
Material : Cast Iron
Condition : New
Price 13500 INR / Piece
Minimum Order Quantity : 1 Piece
Usage : Industrial
Condition : New
Minimum Order Quantity : 1 Unit
Mute : 1
Usage : Industrial
Material : Stainless Steel
Condition : New